Detection of solid and subsolid pulmonary nodules with lung MRI: performance of UTE, T1 gradient-echo, and single-shot T2 fast spin echo
Cancer Imaging volume 23, Article number: 17 (2023)
Abstract
Background
Although MRI is a radiation-free imaging modality, it has historically been limited in lung imaging due to inherent technical restrictions. The aim of this study is to explore the performance of lung MRI in detecting solid and subsolid pulmonary nodules using T1 gradient-echo (GRE) (VIBE, Volumetric interpolated breath-hold examination), ultrashort time echo (UTE) and T2 Fast Spin Echo (HASTE, Half fourier Single-shot Turbo spin-Echo).
Methods
Patients underwent a lung MRI in a 3 T scanner as part of a prospective research project. A baseline Chest CT was obtained as part of their standard of care. Nodules were identified and measured on the baseline CT and categorized according to their density (solid and subsolid) and size (> 4 mm/ ≤ 4 mm). Nodules seen on the baseline CT were classified as present or absent on the different MRI sequences by two thoracic radiologists independently. Interobserver agreement was determined using the simple Kappa coefficient. Paired differences were compared using nonparametric Mann-Whitney U tests. The McNemar test was used to evaluate paired differences in nodule detection between MRI sequences.
Results
Thirty-six patients were prospectively enrolled. One hundred forty-nine nodules (100 solid/49 subsolid) with mean size 10.8 mm (SD = 9.4) were included in the analysis. There was substantial interobserver agreement (k = 0.7, p = 0.05).
Detection for all nodules, solid and subsolid nodules was respectively; UTE: 71.8%/71.0%/73.5%; VIBE: 61.6%/65%/55.1%; HASTE 72.4%/72.2%/72.7%. Detection rate was higher for nodules > 4 mm in all groups: UTE 90.2%/93.4%/85.4%, VIBE 78.4%/88.5%/63.4%, HASTE 89.4%/93.8%/83.8%. Detection of lesions ≤4 mm was low for all sequences. UTE and HASTE performed significantly better than VIBE for detection of all nodules and subsolid nodules (diff = 18.4 and 17.6%, p = < 0.01 and p = 0.03, respectively). There was no significant difference between UTE and HASTE. There were no significant differences amongst MRI sequences for solid nodules.
Conclusions
Lung MRI shows adequate performance for the detection of solid and subsolid pulmonary nodules larger than 4 mm and can serve as a promising radiation-free alternative to CT.
Background
Lung cancer is the second most frequent cancer worldwide and is the leading cause of cancer-related mortality [1]. In the last decades, multiple early lung cancer detection initiatives based on screening at-risk populations with low-dose chest CT have been undertaken with the rationale that early stage diagnosis and subsequent treatment results in a more favorable prognosis. In 2011, the US national lung screening trial (NLST) showed a 20% lung cancer mortality reduction in a high risk population that underwent annual screening with low-dose chest CT as compared to chest x-ray [2]. In 2016, the pilot UK lung cancer screening (UKLS) demonstrated that screening an at-risk population allows for the detection of lung cancer in early stages and that this intervention is cost effective [3].
However, low dose chest CT inherently involves an augmented exposure to medical radiation which increases the long-term risk of radiation-induced cancer [4, 5]. It has been estimated that the lifetime attributable risk of low dose chest CT screening to develop lung cancer ranges from 5.5-8.8 per 10,000 individuals after 10 years, and for other major cancers to 1.4 to 2.6 per 10,000 individuals [6].
Pulmonary imaging with MRI was historically limited by low proton density, rapid signal decay and respiratory motion artifact. However, technical developments achieved by stronger gradients, increased field homogeneity, parallel imaging reconstruction and shorter echo times has made lung MRI an interesting alternative to CT [7, 8].
A free-breathing 3D radial ultrashort echo time (UTE) technique has shown promising results in the assessment of pulmonary disease and specifically in the detection of pulmonary nodules, given that shorter echo times reduce magnetic susceptibility artifacts in the soft tissue-air interfaces [9]. Similarly, single-shot fast-spin-echo sequences have also shown high sensitivity for the detection of pulmonary nodules greater than 5 mm [10]. Clinical and experimental studies have shown a threshold detection size of 3-4 mm for solid nodules and detection rates ranging from 60 to 90% for 5-8 mm lesions and nearly 100% for nodules over 8 mm [7, 11, 12].
The role of lung MRI imaging in lung cancer screening or as a surveillance tool still needs to be defined, and more evidence is needed at this early stage of investigation. Evidence for part solid and ground glass nodules (“adenocarcinoma spectrum lesions”) is particularly sparse. The main objective of the present work is to study the accuracy of T1 gradient-echo (Volumetric interpolated breath-hold examination or VIBE), ultrashort time echo (UTE) and single-shot T2 fast spin echo (Half fourier Single-shot Turbo spin-Echo or HASTE) sequences for the detection of pulmonary nodules. The secondary objective is to determine if there are differences in the detection rate of solid and subsolid nodules.
Methods
Patients
This study was approved by the Research Ethics Board of Sunnybrook Health Sciences Centre in Toronto, Canada, with informed consent obtained from all participants. Patients with Stage I non-small cell lung cancer (NSCLC) eligible for stereotactic body radiotherapy (SBRT) were prospectively enrolled between April 22, 2017 and June 1, 2021 as part of another study. Each patient had a baseline chest CT, which demonstrated the known primary malignancy and a variable number of additional small pulmonary nodules.
Image acquisition
Patients underwent a pre-treatment lung MRI using a 3 T scanner (Magnetom Prisma, Siemens Healthcare) equipped with an 18-element body coil. The protocol included ultra-short time echo (UTE), spoiled 3D T1 gradient echo (VIBE), and single-shot turbo spin echo (HASTE) sequences. The image acquisition parameters for each sequence are summarized in Table 1. Different vendor names for each MRI sequence are highlighted in Table 2.
For UTE and VIBE, the field of view (FOV) included the whole chest, whereas the HASTE FOV was limited to the lung parenchyma adjacent to the known primary malignancy given institutional magnet time restrictions. The HASTE and T1 GRE (VIBE) scans were acquired in the axial plane. Short rectangular pulses, a center-out acquisition and variable TE encoding were applied to achieve an effective TE of 50 μs. To reduce the number of through-plane phase-encoding steps in UTE, images were acquired in the coronal plane [13].
After the completion of lung MRI, all patients had surveillance chest CT as part of their standard-of-care, 3-4 months after SBRT.
Image interpretation
All noncalcified pulmonary nodules were identified and measured on the baseline CT by two fellowship-trained thoracic radiologists in consensus (R1 and R2; 4 and 17 years of experience) and the maximum diameter was recorded. Nodules were categorized into groups according to their density on CT; solid or subsolid (including pure ground-glass and part-solid nodules); and size > 4 mm or ≤ 4 mm, and this evidence was used as the gold standard.
The baseline CTs were done in multiple different centers as patients were referred to our tertiary hospital for treatment. Hence, the CT technical parameters were not explicitly controlled and were performed with or without intravenous contrast and in different scanners (16 to 64-multidetector CTs) across the cohort of participants. Images were reconstructed using a soft tissue and a lung kernel, with a slice thickness ranging between 1.25 to 3 mm. Readers ensured that all baseline examinations were of diagnostic quality and encompassed the whole chest before being included in the study. The nodules were cross-referenced in the follow-up chest CT to ensure that they were present at the time of the lung MRI.
The included pulmonary nodules were classified as present or absent on the different MRI sequences by R1 and R2 independently. Radiologists had access to the baseline CT images while reading the MRI images, allowing for a side-by-side comparison.
Statistical analysis
Detection of nodules for each different group (according to density and size) was expressed in percentages with 95% confidence intervals (CIs). Interobserver agreement was determined using the simple Kappa coefficient.
Continuous variables were reported as means ± standard deviation or as medians and range, as appropriate. Paired differences were compared using nonparametric Mann-Whitney U tests.
The McNemar test was used to evaluate paired differences in nodule detection between MRI imaging techniques. A p-value lower than 0.05 was considered significant. Choice of non-parametric test was to accommodate positively skewed variables. All statistical analyses were carried out using SAS 9.4 software (Cary, NC, USA).
Results
Patients
Patient demographics and characteristics are listed in Table 3. Thirty-six patients, 20 males and 16 females were prospectively enrolled. The mean age of the participants was 75.0 (SD ± 8.3) years. Mean time between the baseline CT and lung MRI was 43.2 ± 13.2 days. The majority of the primary lung cancers were solid (n = 28) with 8 cancers being subsolid. All baseline CTs were classified as providing diagnostic quality, with no major artifacts. The most frequent additional findings were emphysema (n = 19) and minor small airway changes such as tree-in-bud or mucus plugging (n = 9). Four patients had focal post-radiation fibrosis from a previously treated malignancy.
Pulmonary nodules
The nodule characteristics are listed in Table 4. A total of 149 nodules - 100 solid (67.1%) and 49 (32.9%) subsolid - identified on the baseline CT were included in the analysis. Given the limitations of HASTE, only 124 nodules could be potentially visualized on this sequence as some nodules were not included in the FOV. All the nodules were included in the FOV of UTE and VIBE.
The mean diameter of all nodules was 10.8 (SD ± 9.4) mm. There was no significant difference in size between solid and subsolid nodules [10.4 (SD ± 9.6) mm versus 11.5 (SD ± 8.8) mm respectively; p = 0.09]. There was no significant difference between nodules included in the limited FOV of HASTE and the nodules included in UTE and VIBE (VIBE/UTE = 10.8 (SD ± 9.4) mm; HASTE = 11.5 (SD ± 9.9) mm; p = 0.6) (Table 4).
Nodule detection and sequence comparison
Table 5 lists the different detection rates in all sequences regarding the size and density of nodules, as well as detailed comparison among the different MRI sequences.
There was substantial interobserver agreement between the 2 readers (Kappa coefficient = 0.7; 95% CI = 0.62-0.77). Overall nodule detection (solid and subsolid nodules) for UTE was 71.8% (107/149), VIBE 61.7% (92/149) and HASTE 72.4% (89/123). The known primary lung malignancy was depicted by all the different MRI sequences in each case. Nodule detection markedly increased when nodules smaller than 5 mm were excluded from the analysis: UTE = 90.2% (92/102), VIBE = 78.4% (80/102), and HASTE = 89.4% (76/85). Detection was lower for smaller nodules (≤ 4 mm): UTE 31.9% (15/47), VIBE 25.5% (12/47), and HASTE 34.2% (13/38). Detection rate was significantly higher for UTE and HASTE when compared to VIBE for nodules of all size and for nodules larger than 4 mm (UTE vs VIBE: p = < 0.01; HASTE vs VIBE: p = 0.03 for all nodules and < 0.01 for nodules > 4 mm). There were no significant differences for smaller nodules (≤ 4 mm).
Detection of solid nodules was similar among all sequences and ranged from 65.0 - 72.2%. For solid nodules > 4 mm, detection rate was > 85% in all sequences. Detection of smaller solid nodules (≤ 4 mm) ranged from 28.2 - 38.7%. There were no significant differences between the three imaging sequences regarding the solid nodules (Fig. 1). UTE and HASTE performed equally well in the detection of subsolid nodules with rates comparable to the detection of solid nodules (73.5 and 72.7%, respectively), whereas VIBE showed a significantly lower detection rate (55.1%) (Fig. 2). For subsolid nodules > 4 mm, UTE and HASTE showed a detection rate > 80%, whereas VIBE had a significantly lower detection rate of 63.4%. UTE and HASTE performed significantly better than VIBE for subsolid nodules of all sizes and those > 4 mm (UTE vs VIBE: p = < 0.01; HASTE vs VIBE: p = 0.03 for all nodules and < 0.01 for nodules > 4 mm). Detection was low for nodules ≤4 mm in all 3 sequences, ranging from 12.5 - 14.2% and there were no significant differences amongst the 3 sequences. Only one nodule was detected in each sequence in this category.
Discussion
This prospective study demonstrates, that lung MRI has the ability to detect small subsolid and solid nodules larger than 4 mm accurately and in a comparable way to thin section chest CT. Specifically, UTE and HASTE outperformed T1-GRE (VIBE) in all nodules irrespective of size and density, in all subsolid nodules irrespective of size, and in all nodules and subsolid nodules larger than 4 mm. In addition, there was substantial interobserver agreement for pulmonary nodule detection amongst the different MRI sequences.
Although chest CT remains the gold standard imaging modality for the detection of pulmonary nodules, a major caveat is radiation exposure, which is not negligible despite the broad application of low dose techniques and iterative reconstruction [14,15,16]. With the massification of chest CT, small sub-centimeter pulmonary nodules have become a frequent finding. These most often represent an incidental finding and are not clinically significant. There is a non-linear relationship between nodule size and probability of cancer [17]: the probability of cancer of a nodule smaller than 5 mm is approximately 0.4%, whereas it reaches 15.2% for nodules larger than 10 mm [18, 19]. In 2017, the Fleischner society updated their guidelines for the management of incidental pulmonary nodules [20] increasing the threshold of size requiring follow-up to 6 mm, compared to their first version published in 2005 [18] to reduce the number of unnecessary follow-up examinations. Still, many patients will require at least one follow-up CT to confidently exclude malignancy [20].
Significant technical advancements in the last 2 decades have established lung MRI as a complementary and promising radiation-free imaging modality for the detection of pulmonary nodules and thoracic oncology that has the potential to become an attractive alternative to CT.
The overall nodule detection rate of lung MRI in our study ranged between 72% (UTE and HASTE) and 62% (VIBE). Reported rates regarding HASTE are variable in previous studies, with Shroeder et al. reporting an overall nodule detection rate of 85.4% [21], whereas Cieswanowski et al. reported an overall nodule detection rate of 26% [22]. The authors attributed these discrepancies to different hardware and software settings [22]. Many researchers have used 3D T1-weighted GRE (VIBE) with Chandarana et al. reporting an overall nodule detection rate of 62% [23] and Dews et al. reporting a sensitivity of 88% [24]. In a previous study where the authors compared the performance of UTE versus standard and low dose chest CT, they reported a detection rate of 93% for all nodules [25].
The detection rate for all nodules larger than 4 mm in the present study was 89.4% with HASTE, 78.4% with VIBE and 90.2% with UTE. Shroeder et al. reported a detection rate of 86.3% with HASTE for nodules between 3 and 5 mm and 95.7% for nodules between 6 and 10 mm [21]. Cieszanowski et al. reported a rate of 20.3% with HASTE for nodules between 4 and 8 mm and 50% for nodules larger than 8 mm [22]. Dews et al. reported a detection rate of 87.2% with VIBE for nodules between 5 and 10 mm [24]. Olthof et al. reported a low rate of 25% for breath-hold VIBE for nodules between 5 and 10 mm [13], and Yu et al. reported a rate of 62% for breath-hold VIBE for nodules between 4 and 6 mm and 93% for free breathing radial VIBE [26]. Regarding nodules larger than 4 mm, Ohno et al. reported a rate of 74.1% for nodules 4-6 mm - compared to 71.8% in our study – and 94.7% for all solid nodules 6-8 mm – compared to 93.4% in the present study [25]. Ohno et al. did not assess nodules smaller than 4 mm [25].
The detection rate for subsolid nodules irrespective of size was 73.5% for UTE, 72.7% for HASTE and 55.1% for VIBE in the present study. When assessing subsolid nodules larger than 4 mm the detection rates increased to 85.4% for UTE, 83.8% for HASTE and 63.4% for VIBE. Interestingly although UTE performed slightly better than HASTE in the detection of all subsolid nodules and subsolid nodules larger than 4 mm, this did not reach statistical significance. Regarding subsolid nodules larger than 4 mm, Ohno et al. reported a rate of 33.3% with UTE for nodules between 4 and 6 mm and 100% for nodules between 6 and 8 mm [25]. In a study with a lung screening population, 7 out of 8 subsolid nodules were visible on T2-STIR and T2 but there is no information regarding their size [27]. Of note, none of the subsolid nodules in the same study were detected by THRIVE (3D-T1), which is equivalent to VIBE [27].
MRI Sequences with ultra-short TE have been introduced to mitigate the drawbacks of the shorter T2* relaxation times of the lung parenchyma compared to other tissues, secondary to inhomogeneous magnetic susceptibility of lung parenchyma resulting from its air/soft tissue interfaces [28]. Previous studies have shown that the UTE sequence can be used safely to detect significant detail of the lung parenchyma and interstitium comparable to CT including emphysema, interstitial lung disease and airway disease [29,30,31]. The present study showed comparable detection rate of subsolid nodules > 4 mm (85%) when compared to the only other study in the literature assessing subsolid nodule detection, where 81.8% rate was reported for subsolid nodules 6-8 mm [25]. The authors of that study suggested that free-radiation lung MRI imaging with UTE could play a significant role in detection and evaluation of nodule type in lung cancer screening, lung metastasis follow up as well as in the assessment of treatment response.
Studies have shown improved detection rates when using free breathing VIBE resulting in 94% detection rate compared to breath-hold VIBE (64.3%), however the scanning time is significantly increased from 14 sec to 7 min which might not be clinically applicable [13, 26]. In the present study, breath-hold VIBE had the lowest detection rate compared to UTE and HASTE.
HASTE is a single-shot turbo spin echo T2 weighted sequence and has been studied extensively for the detection of pulmonary nodules with comparable results to the present study [21, 32, 33], especially for nodules larger than 5 mm, however it has not been studied for detection of subsolid pulmonary nodules. In a study by Meier-Shroers et al., T2 weighted imaging with an acquisition time of 3.18 min. Was most reliable for the detection of pulmonary nodules ≥6 mm, however they did not evaluate HASTE [27].
There were limitations to this study. First, the number of patients was small, although the total number of nodules was relatively large when compared to other series [10, 11, 27], and smaller nodules were not excluded. Second, radiologists had access to the CT images while reading the MRI sequences, which may have caused bias towards overcalling pulmonary nodules on the MRI images. This could be assessed in future studies by blinding the readers to the CT images, simulating real-life clinical conditions. Third, there was a time interval between the baseline CT and lung MRI (43.2 ± 13.2 days). It is possible that some lesions might have slightly increased in size in the interval, although this is likely non-significant as the average doubling time (growth-rate) of malignant lesions is longer than this period, and benign lesions tend to remain stable [34, 35]. Fourth, as HASTE did not include the whole chest, some nodules were excluded from the analysis for this sequence in particular. Fifth, we did not assess morphological characteristics of the nodules such as shape, border, location or calcification. Lastly, a number of nodules did not have histologic confirmation and therefore we did not attempt to differentiate malignant from benign nodules.
Conclusion
In conclusion, this study has shown that clinically significant solid and subsolid nodules larger than 4 mm can be accurately detected by lung MRI using breath-hold UTE and HASTE that could be completed in less than 5 min of magnet time [12]. Fast non-contrast lung MRI is gaining its role in the routine clinical practice as a radiation-free alternative to CT for nodule detection and could be used in lung cancer screening, follow-up of pulmonary nodules, and in the metastatic work-up of oncologic patients.
Abbreviations
- GRE:
-
Gradient-echo
- VIBE:
-
Volumetric interpolated breath-hold examination
- UTE:
-
Ultrashort time echo
- MRI:
-
Magnetic Resonance
- HASTE:
-
Half fourier Single-shot Turbo spin-Echo
- CT:
-
Computed Tomography
- NLST:
-
National Lung Screening Trial
- UKLS:
-
United Kingdom lung cancer screening
- NSCLC:
-
Non-small cell lung cancer
- SBRT:
-
Stereotactic body radiotherapy
- FOV:
-
Field of view
References
World Health Organization. All cancers 2020, https://gco.iarc.fr/today/data/factsheets/cancers/39-All-cancers-fact-sheet.pdf. Accessed 12 Oct 2022.
The National Lung Screening Trial Research Team. Reduced lung-cancer mortality with low-dose computed tomographic screening. N Engl J Med. 2011;365:395–409.
Field JK, Duffy SW, Baldwin DR, Whynes DK, Devaraj A, Brain KE, et al. UK lung cancer RCT pilot screening trial: baseline findings from the screening arm provide evidence for the potential implementation of lung cancer screening. Thorax. 2016;71:161–70.
Shah DJ, Sachs RK, Wilson DJ. Radiation-induced cancer: a modern view. Br J Radiol. 2012;85(1020):e1166–73.
Larke FJ, Kruger RL, Cagnon CH, Flynn MJ, McNitt-Gray MM, Wu X, et al. Estimated radiation dose associated with low-dose chest CT of average-size participants in the National Lung Screening Trial. AJR Am J Roentgenol. 2011;197:1165–9.
Rampinelli C, De Marco P, Origgi D, Maisonneuve P, Casiraghi M, Veronesi G, et al. Exposure to low dose computed tomography for lung cancer screening and risk of cancer: secondary analysis of trial data and risk-benefit analysis. BMJ. 2017;356:j347.
Biederer J, Ohno Y, Hatabu H, Schiebler ML, van Beek EJR, Vogel-Claussen J, et al. Screening for lung cancer: Does MRII have a role? Eur J Radiol. 2017;86:353–60.
Sieren JC, Ohno Y, Koyama H, Sugimura K, McLennan G. Recent technological and application developments in computed tomography and magnetic resonance imaging for improved pulmonary nodule detection and lung cancer staging. J Magn Reson Imaging. 2010;32:1353–69.
Ohno Y, Koyama H, Yoshikawa T, Seki S, Takenaka D, Yui M, et al. Pulmonary high-resolution ultrashort TE MRI imaging: Comparison with thin-section standard- and low-dose computed tomography for the assessment of pulmonary parenchyma diseases. J Magn Reson Imaging. 2016;43:512–32.
Schroeder T, Ruehm SG, Debatin JF, Ladd ME, Barkhausen J, Goehde SC. Detection of Pulmonary Nodules Using a 2D HASTE MRI Sequence: Comparison with MDCT. Am J Roentgenol. 2005;185:979–84.
Burris NS, Johnson KM, Larson PE, Hope MD, Nagle SK, Behr SC, et al. Detection of small pulmonary nodules with ultrashort echo time sequences in oncology patients by using a PET/MRI system. Radiology. 2016;278:239–46.
Koyama H, Ohno Y, Kono A, Takenaka D, Maniwa Y, Nishimura Y, et al. Quantitative and qualitative assessment of non-contrast-enhanced pulmonary MRI imaging for management of pulmonary nodules in 161 subjects. Eur Radiol. 2008;18:2120–31.
Olthof SC, Reinert C, Nikolaou K, et al. Detection of lung lesions in breath-hold VIBE and free-breathing Spiral VIBE MRII compared to CT. Insights. Imaging. 2021:175. https://doi.org/10.1186/s13244-021-01124-0.
Rubin GD. Lung nodule and cancer detection in computed tomography screening. J Thorac Imaging. 2015;30:130–8.
Linet MS, Slovis TL, Miller DL, Kleinerman R, Lee C, Rajaraman P, et al. Cancer risks associated with external radiation from diagnostic imaging procedures. CA Cancer J Clin. 2012;62:75–100.
Fujii K, McMillan K, Bostani M, Cagnon C, McNitt-Gray M. Patient size–specific analysis of dose indexes from CT lung cancer screening. Am J Roentgenol. 2016;208:144–9.
McWilliams A, Tammemagi MC, Mayo JR, Roberts H, Liu G, Soghrati K, et al. Probability of cancer in pulmonary nodules detected on first screening CT. N Engl J Med. 2013;369:910–9.
Horeweg N, van Rosmalen J, Heuvelmans MA, van der Aalst CM, Vliegenthart R, Scholten ET, et al. Lung cancer probability in patients with CT-detected pulmonary nodules: a prespecified analysis of data from the NELSON trial of low-dose CT screening. Lancet Oncol. 2014;15:1332–41.
Wahidi MM, Govert JA, Goudar RK, Gould MK, McCrory DC. Evidence for the treatment of patients with pulmonary nodules: when is it lung cancer?: ACCP evidence-based clinical practice guidelines (2nd edition). Chest. 2007;132(Suppl 3):94s–107s.
MacMahon H, Naidich DP, Goo JM, Lee KS, Leung ANC, Mayo JR, et al. Guidelines for management of incidental pulmonary nodules detected on CT images: from the Fleischner Society 2017. Radiology. 2017;284:228–43.
Schroeder T, Ruehm SG, Debatin JF, Ladd ME, Barkhausen J, Goehde SC. Detection of pulmonary nodules using a 2D HASTE MRI sequence: comparison with MDCT. AJR Am J Roentgenol. 2005;185:979–84.
Cieszanowski A, Lisowska A, Dabrowska M, Korczynski P, Zukowska M, Grudzinski IP, et al. MRI imaging of pulmonary nodules: detection rate and accuracy of size estimation in comparison to computed tomography. PLoS One. 2016. https://doi.org/10.1371/journal.pone.0156272.
Chandarana H, Heacock L, Rakheja R, DeMello LR, Bonavita J, Block TK, et al. Pulmonary nodules in patients with primary malignancy: comparison of hybrid PET/MRI and PET/CT imaging. Radiology. 2013;268:874–81.
Dewes P, Frellesen C, Al-Butmeh F, Albrecht MH, Scholtz JE, Metzger SC, et al. Comparative evaluation of non-contrast CAIPIRINHA-VIBE 3T-MRII and multidetector CT for detection of pulmonary nodules: In vivo evaluation of diagnostic accuracy and image quality. Eur J Radiol. 2016;85:193–8.
Ohno Y, Koyama H, Yoshikawa T, Kishida Y, Seki S, Takenaka D, et al. Standard-, reduced-, and no-dose thin-section radiologic examinations: comparison of capability for nodule detection and nodule type assessment in patients suspected of having pulmonary nodules. Radiology. 2017;284:562–73.
Yu N, Yang C, Ma G, Dang S, Ren Z, Wang S, et al. Feasibility of pulmonary MRII for nodule detection in comparison to computed tomography. BMC Med Imaging. 2020. https://doi.org/10.1186/s12880-020-00451-w.
Meier-Schroers M, Homsi R, Schild HH, Thomas D. Lung cancer screening with MRI: characterization of nodules with different non-enhanced MRII sequences. Acta Radiol. 2019;60:168–76.
Ciliberto M, Kishida Y, Seki S, Yoshikawa T, Ohno Y. Update of MRI Imaging for Evaluation of Lung Cancer. Radiol Clin N Am. 2018;56:437–69.
Ohno Y, Koyama H, Yoshikawa T, et al. Pulmonary high-resolution ultrashort TE MRI imaging: comparison with thin-section standard- and low-dose computed tomography for the assessment of pulmonary parenchyma diseases. J Magn Reson Imaging. 2016;43:512–32.
Ohno Y, Koyama H, Yoshikawa T, et al. T2* measurements of 3-T MRI with ultrashort TEs: capabilities of pulmonary function assessment and clinical stage classification in smokers. AJR Am J Roentgenol. 2011;197:W279–85.
Ma W, Sheikh K, Svenningsen S, et al. Ultra-short echo-time pulmonary MRI: evaluation and reproducibility in COPD subjects with and without bronchiectasis. J Magn Reson Imaging. 2015;41:1465–74.
Vogt FM, Herborn CU, Hunold P, et al. HASTE MRI versus chest radiography in the detection of pulmonary nodules: comparison with MDCT. AJR Am J Roentgenol. 2004;183:71–8.
Regier M, Kandel S, Kaul MG, et al. Detection of small pulmonary nodules in high-field MRI at 3 T: evaluation of different pulse sequences using porcine lung explants. Eur Radiol. 2007;17:1341–51.
Larici AR, Farchione A, Franchi P, Ciliberto M, Cicchetti G, Calandriello L, et al. Lung nodules: size still matters. Eur Respir Rev. 2017;26:170025.
Yankelevitz DF, Gupta R, Zhao B, Henschke CI. Small pulmonary nodules: evaluation with repeat CT-preliminary experience. Radiology. 1999;212:561–6.
Acknowledgements
Not applicable.
Funding
This study has been funded by the AFP 2016-2017 Sunnybrook Innovation Grant.
Author information
Authors and Affiliations
Contributions
FS: Selection of clinical data, reader for the pulmonary nodules on MRI and CT, statistical analysis, interpretation of results, drafting of the paper, revision of the paper. PT: Statistical analysis, revision of the paper. CH: MRI sequence protocols, interpretation of results, revision of the paper. PC: Collection of clinical data, revision of the paper. IP: Collection of clinical data, revision of the paper. YU: Collection of clinical data, revision of the paper. AL: Collection of clinical data, revision of the paper. MT: Collection of clinical data, revision of the paper. SG: MRI sequence protocols, interpretation of results, revision of the paper. AO: Conception and study design, reader for the pulmonary nodules on MRI and CT, collection of clinical data, interpretation of results, drafting of the paper, revision of the paper. The authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study was approved by the Research Ethics Boards of our institution (REB project identification number 039-2017). All methods were carried out in accordance with relevant guidelines and regulations.
Consent for publication
An informed consent was obtained from all participants.
Competing interests
The authors declare that they have no competing interests.
Additional information
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
About this article
Cite this article
Sanchez, F., Tyrrell, P.N., Cheung, P. et al. Detection of solid and subsolid pulmonary nodules with lung MRI: performance of UTE, T1 gradient-echo, and single-shot T2 fast spin echo. Cancer Imaging 23, 17 (2023). https://doi.org/10.1186/s40644-023-00531-4
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s40644-023-00531-4