Skip to main content

Table 1 Detailed sequence parameters of WB-DWI/MRI (whole body diffusion-weighted magnetic resonance imaging)

From: Whole-body diffusion-weighted MRI for operability assessment in patients with colorectal cancer and peritoneal metastases

 

DWI

T2 SSTSE

Contrast-enhanced 3D T1 gradient-echo

Transverse

Coronal

Sagittal

Coronal

Transverse

Coronal

Transverse

Image stations head to mid-thigh

4

MPR

MPR

3

abdominopelvic (2)

abdominopelvic (2)

Chest (1)

Respiration

Free breathing

  

respiratory

15 s breath-hold

Fat suppression

STIR (TI = 250 ms)

  

none

SPAIR (mDIXON)

SPAIR (mDIXON)

SPAIR (eTHRIVE)

b-values (s/mm2)

0–1000

  

none

none

none

none

Parallel imaging factor

2.5

  

4

2

2

2

Repetition time (TR) (ms)

8454

  

3000

3.6

3.6

3.2

Echo time (TE) (ms)

67

  

87

1.25–2.20

1.25–2.20

1.5

Slice thickness (mm)

5

5

5

6

2.5

2.5

2.5

Slice number

50/station

  

35/station

90

133

148

Intersection gap (mm)

0.1

  

0.6

0

0

0

Field of view (FOW) (mm)

420 × 329

  

375 × 447

375 × 304

400 × 352

375 × 304

Acquired voxel size (mm)

4.57 × 4.71

  

1 × 1

1.49 × 1.5

1.49 × 1.5

1.49 × 1.5

Reconstructed voxel size (mm)

2.19 × 2.16

  

0.93 × 0.93

0.71 × 0.71

0.71 × 0.71

0.98 × 0.97

Number of signal averages (NSA)

1

  

1

1

1

1

  1. DWI diffusion-weighted imaging, mDIXON multi-echo 2-point Dixon, eTHRIVE T1-weighted high-resolution isotropic volume examination, SSTSE single-shot turbo spin-echo imaging, STIR short T1 inversion recovery, SPAIR spectrally adiabatic inversion recovery