Skip to main content

Table 1 MR sequence parameters

From: Automated deep-learning system in the assessment of MRI-visible prostate cancer: comparison of advanced zoomed diffusion-weighted imaging and conventional technique

Parameter

T2-weighted imaging

f-DWI

z-DWI

Sequence

TSE

SS-EPI

zoomed SS-EPI

Field of view (mm2)

200 × 200

380 × 280

190 × 109

Scan matrix

320 × 256

178 × 132

100 × 56

Voxel size (mm3)

0.6 × 0.6 × 3.5

2.1 × 2.1 × 3.0

1.9 × 1.9 × 3.0

PE Dir.

Left- > Right

Anterior- > Posterior

Left- > Right

Echo time (ms)

101

79

76

Repetition time (ms)

6000

4200

3800

Bandwidth (Hz/px)

200

1872

1612

b-value (s/mm2)

NA

50, 1000, 1500

NEX

2

1, 3, 6

Fat suppression

NA

FatSat

Acceleration factor

2 (GRAPPA)

2 (GRAPPA)

NA

Acquisition time (min)

2:08

3:05

2:37

  1. PE Dir. Phase Encoding direction, GRAPPA Generalized auto-calibrating partial parallel acquisition, NA not applicable