Table 5 Correlation analysis of DLP (mGy-cm) and patient diameter

From: Chest-abdomen-pelvis CT for staging in cancer patients: dose effectiveness and image quality using automated attenuation-based tube potential selection

  

Patients

DLP (mGy · cm)

Body diameter (mm): lateral

Body diameter (mm): transverse

Spearman correlation coefficient (rho) – lateral diameter/DLP

Spearman correlation coefficient (rho) – transverse diameter/DLP

Group 1

Patients 100 kV

32

564 (390–871)

327.3 (280.7–408)

228.2 (150.7–317.2)

0.81

0.87

Patients 120 kV

17

833 (470–1235)

351.2 (281.1–452.5)

249 (165.7–346.8)

0.80

0.85

Patients 140 kV

6

1302 (1157–2203)

442.5 (401.5–478)

290.1 (259–409)

0.42

0.43

Group 2

Patients 120 kV

55

756 (345–2267)

330.8 (196.3–496.1)

241.9 (155.9–446.3)

0.88

0.82